KELIBRON Guides

Plate inspection & GD&T

Surface plate and height gauge

3 min readUpdated

Key answer

To measure a step by comparison, clean the plate and the contact faces, set or note the reading on the nominal gauge-block stack, then slide the height-gauge base to the step. Add the signed reading difference to the stack size; keep the base in contact with the plate throughout.

On this page
  1. Measure a step height by comparison
  2. Good habits on the plate
  3. The plate is the reference
  4. The accessories that pose the part
  5. In KELIBRON
  6. References
  7. Practise the shared basics on a caliper
  8. Questions

Measure a step height by comparison

Wipe the plate, the base and the workpiece underside clean and remove burrs before setting up. Do not hammer on the plate or drag work across it. Protect it during storage as appropriate for its material; metal surfaces that can rust need corrosion protection.

The measurement is comparative, like comparison against gauge blocks: bring the scribing jaw down onto a stack of gauge blocks of the nominal size and set the zero (or note the reading), slide the base to the measured point and read again. Height = nominal size + (reading − zero reading).

Slide the base without lifting it. Losing contact with the plate loses the reference for the height comparison.

StepReading
Zero on 30 mm gauge blocks0.00 mm
Step face+0.12 mm
Step height30 + 0.12 = 30.12 mm

Good habits on the plate

Clean first: a burr under the base or the workpiece tilts the whole setup by its own height. Move gauges on their bases, not by their beams. Finish a comparative setup by rocking the contact to find the true extreme, the same habit the gauge-block comparison scene trains.

A scribing jaw relies on your eye against a line; for tighter work clamp a lever indicator to the slider, which reads where the needle stops instead of where the eye guesses.

The plate is the reference

A surface plate is a thick, stiff slab whose working face is finished flat enough to serve as the reference plane of the bench. Cast-iron plates follow the product standard GB/T 22095-2008; both cast-iron and granite plates are verified to JJG 117-2013《平板检定规程》(which superseded JJG 117-2005), covering first verification, subsequent verification and in-service checks.

Plates are graded by accuracy — 0, 1, 2 and 3, the smaller the number the flatter the face. The flatness limit values are read from JJG 117-2013; until the standard text is at hand this article quotes the grade system only, not the table.

A height gauge (height calliper, GB/T 21390-2008《游标、带表和数显高度卡尺》) is a beam on a heavy base: the slider carries a scribing jaw or a clamp for a lever indicator, and reads on a vernier, a dial or a display. Its measuring datum is the underside of the base — the beam only carries the slider.

The accessories that pose the part

A dial stand clamps an indicator rigidly to the plate and parks it exactly where the reading happens. A V-block seats a cylinder on its axis. A square box (angle plate) with a V-groove and square faces holds a part upright so a face becomes measurable. A try square checks a vertical direction against the plate.

None of them gives a reading; all of them decide what the reading means. Form and position checks on the plate start with posing the part, and the posing is where the datum actually comes from — see datums.

In KELIBRON

KELIBRON has indicator stands, but no surface plate or height-gauge exercise. The existing comparison scene can train zeroing and reading a difference, not a complete height-gauge inspection. The comparative habit this lesson needs is already practised: the gauge-block-30-compare scene walks the same zero-then-read difference on an existing bench. Flatness checks on the plate follow in flatness and straightness.

References

  1. JJG 117-2013《平板检定规程》(代替 JJG 117-2005) (cited by number; no public page) · 国家质量监督检验检疫总局
  2. GB/T 22095-2008《铸铁平板》 · 国家质量监督检验检疫总局、国家标准化管理委员会
  3. GB/T 21390-2008《游标、带表和数显高度卡尺》 · 国家质量监督检验检疫总局、国家标准化管理委员会
  4. 金莹 主编,《公差配合与技术测量》,清华大学出版社,2014(2024 年第 4 次印刷),ISBN 978-7-302-37145-8 (cited by number; no public page) · 清华大学出版社

How the guides are checked against these sources: About KELIBRON.

Practise the shared basics on a caliper

Questions

Why is a surface plate used as the measuring datum?

Because everything on the bench shares its face: workpiece, gauge stand and gauge base all rest on the same plane, so heights and positions have one common reference instead of several private ones.

Do I lift the height gauge to move it?

No. Slide the base along the plate. Lifting loses contact with the reference plane and the reading stops being a height.

How do I measure a step height with a height gauge?

Zero the slider on a gauge block stack of the nominal step height, slide to the step face, and add the signed reading difference: 30 mm stack, +0.12 mm → 30.12 mm. See comparative measurement for the same zero-then-read method.

Practise the basics first

Cleaning, zero checks, square positioning, light even force and reading to the scale interval are the same on every instrument; practise them in a caliper scene.

Open in the lab →