Instrument care
Caliper care and calibration
Before each use: clean the faces, close the jaws and check zero. Calibrate at the interval your quality system sets from how often the caliper is used and the errors found at earlier calibrations; NPL’s good practice guide also suggests a periodic inspection at least once a year. To calibrate, check the caliper against gauge blocks at five or more points over its range and record the result.
On this page
Before every use
Never measure a turning part. Stop the machine before making contact, then clean the caliper, check its movement and zero, and confirm its calibration status.
- Wipe the measuring faces and the part; remove burrs and chips.
- Slide the jaw along the beam: it should move smoothly with no play and no sticking.
- Close the jaws and hold them up to a light: no light gap passes this daily face check, not the whole accuracy assessment; a visible gap needs further inspection for wear.
- Check zero: the vernier zero lines up, the dial needle sits on 0, or the digital display reads 0.00 (if resetting is needed, follow the model’s manual; KELIBRON uses ORIGIN with the jaws closed).
- Check the calibration label is in date.
Calibration with gauge blocks
- Outside jaws: measure gauge blocks at five or more roughly equally spaced points covering the range (NPL GPG 40), and at two sizes, one near zero and one near full range, place the block near the jaw root and near the tips to check that the faces are parallel.
- Inside jaws: a setting ring or a gauge-block stack with jaws.
- Depth rod: gauge blocks wrung to a surface plate.
- Compare each error with the maximum permissible error of the caliper (about ±0.02 to ±0.03 mm for a 150 mm caliper with a resolution or scale interval of 0.01 or 0.02 mm; take the value from the specification or standard).
Standards: ISO 13385-1 defines the design and metrological characteristics of calipers; many countries publish their own verification procedures (in China the verification regulation JJG 30). Comparing errors with the MPE to decide pass or fail is verification or an in-house check; a calibration certificate reports the errors and their uncertainty. Record the date, the results and the next due date. Full procedure, the micrometer and the record: checking a caliper and a micrometer with gauge blocks.
Storage and handling
Store clean and dry in its case, faces about 2 mm apart and the slider not clamped (NPL GPG 40).
For a large caliper that is seldom used, a rust preventive on the sliding and measuring faces helps; avoid rust preventives that leave a coating, which can affect a dial caliper’s calibration. Keep it away from magnets, coolant and heat. Never use a caliper as a scriber, hammer or wrench. Take the battery out of a digital caliper for long storage.
References
- ABS Digimatic Caliper — User’s Manual, No. 99MFA071A · Mitutoyo
- ISO 13385-1:2019 — Geometrical product specifications (GPS) — Dimensional measuring equipment — Part 1: Design and metrological characteristics of callipers · ISO
- JJG 30-2012《通用卡尺》检定规程 (cited by number; no public page) · 国家质量监督检验检疫总局
- Measurement Good Practice Guide No. 40 — Callipers and Micrometers, Issue 2 · National Physical Laboratory
- ISO 10012 — Quality management — Requirements for measurement management systems · ISO
- JCGM 200:2012 — International vocabulary of metrology — Basic and general concepts and associated terms (VIM, 3rd edition) · JCGM / BIPM
How the guides are checked against these sources: About KELIBRON.
Watch it done
Played back in the lab from domain commands; the reading and any verdict come from the measurement model.
Practise in the lab
Questions
How often should a caliper be calibrated?
Your quality system sets the interval from how often the caliper is used and the errors found at earlier calibrations; NPL’s good practice guide also suggests a periodic inspection at least once a year. Check zero before every use regardless.
Can I calibrate a caliper myself?
You can check it with traceable gauge blocks and record the results. Formal calibration is done to a written procedure by an accredited lab or a competent in-house metrology department; where verification is a legal requirement (in China, JJG 30), an authorised metrology body does it.
Try it yourself
Open the same setup in the 3D lab and take the reading yourself.
Related guides
Reading instrumentsMicrometer zero check and zero error correctionCheck a micrometer against its reference before measuring: close the clean faces with the ratchet on a 0–25 mm model, or use the matching setting standard on larger ranges.
Instrument careHow to read a calibration certificateFirst match the certificate ID and range to the instrument you will use.
Instrument careGauge management and calibration recordsEvery instrument used to accept product needs a unique number, a register entry, a calibration interval, a visible status and a record of each calibration.