Reading instruments
How to read and use a caliper
Check zero with the jaws closed, hold the part deep in the jaws and square to the beam, close with light, even force, then read: on a vernier, the main-scale millimetres plus the coinciding line × 0.02 mm; on a dial, millimetres plus the needle’s hundredths; on a digital caliper, the display.
On this page
- Checking the zero
- Measuring an outside diameter
- How to read a vernier caliper (0.02 mm)
- How to read a dial caliper (0.01 mm)
- How to use a digital caliper
- Measuring an inside diameter (bore)
- Measuring a depth
- Parts of a caliper
- Common wrong use and how to correct it
- Judging a part against its tolerance
- References
- Watch it done
- Practise in the lab
- Questions
Checking the zero
Check zero with the caliper off the part: wipe the measuring faces and close the jaws, the two parts that grip the workpiece. The depth rod draws in flush. The reading should be 0.00.
If a digital caliper needs its zero reset, follow the model’s manual; KELIBRON uses ORIGIN with closed jaws, but names and presses vary by maker. On a dial caliper, turn the bezel to zero. On a vernier caliper, note the zero error and subtract it, with its sign, from every reading.
Check before every series of measurements. Measure at room temperature: drawing sizes refer to 20 °C (ISO 1), so a part still warm from machining reads large.
Measuring an outside diameter
- Check the zero with the jaws closed.
- Open the jaws a little wider than the part.
- Put the part deep in the jaws, close to the beam (less Abbe error), with the beam square to the part axis.
- Close the jaws gently with light, even force until both faces touch; read the value.
How to read a vernier caliper (0.02 mm)
Read a 0.02 mm vernier caliper by taking the whole millimetres on the main scale, then adding the aligned vernier line’s number × 0.02 mm. The example below shows both parts of the reading.
- Read the whole millimetres on the main scale just to the left of the vernier zero line.
- Find the one vernier line that lines up exactly with a main-scale line.
- Multiply its number by 0.02 mm and add it to the millimetres.
Example: the vernier zero is just past 40 on the main scale and the 13th vernier line lines up: 40 + 13 × 0.02 = 40.26 mm. The figures on the vernier (0, 1 … 9, and 0 or 10 at the last line, depending on the maker) mark every fifth line (0.1 mm steps).
Look straight at the scales. Seen at an angle the vernier lines appear shifted (parallax) and you can read the wrong line. In KELIBRON the “face the readout” view turns the camera square to the scales.
On a 0.02 mm vernier caliper the vernier plate carries 50 divisions spread over 49 mm, so each vernier division is 0.98 mm, 0.02 mm shorter than a main-scale millimetre.
How to read a dial caliper (0.01 mm)
A dial caliper shows whole millimetres on the beam scale and hundredths on the dial: on most metric dial calipers one revolution of the needle is 1 mm, divided into 100 divisions of 0.01 mm (read the value printed on the dial).
Example: the slider edge is past 28 mm on the beam and the needle points at 3: 28 + 3 × 0.01 = 28.03 mm. Before measuring, close the jaws and turn the bezel so that 0 sits under the needle, then lock the bezel.
How to use a digital caliper
Read the digital display once the jaws make light, even contact with the part. Check zero, the mm / inch unit and whether you are in absolute or INC mode before measuring; a common resolution is 0.01 mm.
KELIBRON sets the absolute origin with ORIGIN while the jaws are closed. For a real instrument, check zero first and reset only as needed, following its manual. ZERO/ABS sets a relative zero for comparison with a master (INC) and returns to absolute; some models use a long press to return. UNIT switches mm / inch, and HOLD freezes the reading, not the jaws.
Measuring an inside diameter (bore)
- Close the jaws to a little less than the bore and put the inside jaws into it.
- Open the jaws until both blades touch the wall.
- Keep them on the diameter: rock gently across the bore for the largest reading, and along the bore axis for the smallest.
- Read the value.
Measuring a depth
- Stand the caliper upright and rest the end of the beam flat on the reference face (the open face of the hole).
- Hold the beam end down and push the depth rod out until it touches the bottom.
- Keep the beam square to the face, then read the value.
Parts of a caliper
A caliper has a beam with the main scale and a fixed jaw, and a slider (frame) that carries the moving jaw and the reading device: a vernier plate, a dial or a digital display.
The large lower jaws are the outside jaws for shafts and other outside sizes. The small upper jaws are the inside jaws for bores and slots. The depth rod runs out of the end of the beam as the jaws open, for depths and steps. The lock screw on top clamps the slider; the thumb rest or thumbwheel moves it.
Common wrong use and how to correct it
| Task | Wrong use | Effect on the reading | Correction |
|---|---|---|---|
| Outside | Caliper tilted along the part axis | Too large (the jaws span an ellipse) | Keep the beam square to the axis |
| Outside | Zero not checked | Off by the zero error | Close the jaws and set the zero first |
| Outside | Chip or dirt on a jaw face | Too large | Wipe the faces and the part |
| Outside | Too much measuring force | Usually too small (the jaws spring; on a real caliper it also depends on slider play) | Close with light, even force |
| Inside | Blades off the bore’s diameter | Too small (a chord) | Rock across the bore for the largest reading |
| Inside | Tilted along the bore axis | Too large (about D / cos α) | Square to the axis; rock along it for the smallest reading |
| Inside | Chip on a blade edge | Too small | Wipe the blades and the bore |
| Depth | Beam end not resting on the face | Too large by the gap | Hold the beam end down first |
| Depth | Caliper not square to the face | Too large (about h / cos α) | Keep the beam perpendicular, whole end face down |
| Depth | Chip on the bottom of the hole | Too small | Clean the hole and the rod tip |
Judging a part against its tolerance
A drawing size such as Ø20 0/−0.05 gives the limits 19.95 mm and 20.00 mm. A part passes when the reading lies within the limits and fails low or high outside them. When the reading is within half a scale interval (or digital step) of a limit, or the tolerance band is narrower than two of those steps, KELIBRON gives no verdict instead. A reading taken with a wrong setup (tilt, unchecked zero, debris, too much force) must never be accepted. KELIBRON gives no verdict in its wrong-use demonstrations or when the caliper is tilted beyond the allowed angle; a slight tilt within that angle is still judged, so keep the caliper square.
KELIBRON is a teaching simulation: it is not a calibrated instrument and its readings are not certified measurement results.
References
- ABS Digimatic Caliper — User’s Manual, No. 99MFA071A · Mitutoyo
- ISO 13385-1:2019 — Geometrical product specifications (GPS) — Dimensional measuring equipment — Part 1: Design and metrological characteristics of callipers · ISO
- JJG 30-2012《通用卡尺》检定规程 (cited by number; no public page) · 国家质量监督检验检疫总局
- ISO 1:2022 — Geometrical product specifications (GPS) — Standard reference temperature for the specification of geometrical and dimensional properties · ISO
- GB/T 21389-2008《游标、带表和数显卡尺》 · 国家质量监督检验检疫总局、国家标准化管理委员会
- Measurement Good Practice Guide No. 40 — Callipers and Micrometers, Issue 2 · National Physical Laboratory
- 南昌大学物理实验中心,《密度的测量》实验讲义(游标卡尺读数;螺旋测微器读数与零点误差修正) · 南昌大学
- T. Doiron, “20 °C — A Short History of the Standard Reference Temperature for Industrial Dimensional Measurements”, J. Res. NIST 112(1), 2007 · NIST
How the guides are checked against these sources: About KELIBRON.
Watch it done
Played back in the lab from domain commands; the reading and any verdict come from the measurement model.
Practise in the lab
Vernier caliperRead a vernier caliper on a Ø25 shaft (readout view)Open in the lab →
Dial caliperRead a dial caliper on a bushing (readout view)Open in the lab →
Digital caliperDigital caliper on a stepped shaft (readout view)Open in the lab →
Digital caliperMeasure the bore of a bushingOpen in the lab →
Digital caliperMeasure the depth of a blind boreOpen in the lab →
Questions
How do I read a 0.02 mm vernier caliper?
Whole millimetres from the main scale left of the vernier zero, plus the number of the vernier line that lines up × 0.02 mm. Example: 40 + 13 × 0.02 = 40.26 mm.
Why does a tilted caliper read too large?
Tilted along the part axis, the jaws span an ellipse that is wider than the diameter; the reading is about L / cos α, so the error grows with the angle.
Try it yourself
Open the same setup in the 3D lab and take the reading yourself.
Related guides
Measurement basicsRecording readingsWrite a reading with every digit the instrument resolves, trailing zeros included, and with its unit.
Reading instrumentsHow to read a vernier caliperTo read a vernier caliper, take the whole millimetres just left of the vernier zero, then add the aligned line’s number times the least count.
Reading instrumentsHow to read a dial caliperReading = millimetres on the beam scale (at the slider’s reading edge) + needle divisions × 0.01 mm.