Measurement uncertainty basics

2 min readUpdated Course: Measurement data, uncertainty and decisions · Lesson 2

Key answer

Error is the measured value minus a reference value (VIM 2.16); uncertainty is a non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used (VIM 2.26). A result is written as value ± U (k = 2, about 95 %).

On this page
  1. Worked example: a caliper reading
  2. Type A and Type B
  3. Why it matters for pass/fail
  4. References
  5. Watch it done
  6. Practise in the lab
  7. Questions
Lab clipDigital caliper: Ø12 ±0.03 pinOpen in the lab →

Worked example: a caliper reading

SourceValueStandard uncertainty
Caliper MPE±0.02 mm0.02/√3 = 0.0115 mm
Resolution0.01 mm0.01/(2√3) = 0.0029 mm
Repeatability (10 readings)s = 0.008 mm0.008/√10 = 0.0025 mm

uc = √(0.0115² + 0.0029² + 0.0025²) ≈ 0.0122 mm (from the unrounded components), U = 2 × 0.0122 = 0.0244, rounded up to 0.025 mm. Mean reading 25.03 mm → 25.030 mm ± 0.025 mm (k = 2). Temperature, force and zero setting add further terms in a full budget. As an introductory budget the three terms above are treated as independent; a rigorous budget first checks for overlap, because calibration already covers part of what the MPE and repeatability terms describe.

Type A and Type B

  • Type A: from repeated readings, using the standard deviation s of n readings; u = s/√n for the mean.
  • Type B: from other information, such as a specification limit ±a with rectangular distribution gives u = a/√3; a resolution δ gives u = δ/(2√3).
  • Combined: uc = √(u₁² + u₂² + …) for independent sources.
  • Expanded: U = k · uc, usually k = 2.

Why it matters for pass/fail

If the tolerance band is 0.05 mm and U is 0.025 mm, half of the band is uncertainty, so the instrument is unsuitable (see the 10:1 rule). By the default rule of ISO 14253-1:2017 (when the supplier proves conformity), conformity is accepted only when the result lies inside the limits reduced on each side by a guard band of about 0.83U (1.65 combined standard uncertainties, clause 5.2.2, with U = 2u and a normal distribution whose standard deviation is clearly smaller than the tolerance; 95 % conformance probability); other decision rules can be agreed.

References

  1. JCGM 100:2008 — Evaluation of measurement data — Guide to the expression of uncertainty in measurement (GUM) · JCGM / BIPM
  2. JCGM 200:2012 — International vocabulary of metrology — Basic and general concepts and associated terms (VIM, 3rd edition) · JCGM / BIPM
  3. JJF 1059.1-2012《测量不确定度评定与表示》 (cited by number; no public page) · 国家质量监督检验检疫总局
  4. ISO 14253-1:2017 — GPS — Inspection by measurement of workpieces and measuring equipment — Part 1: Decision rules for verifying conformity or nonconformity with specifications · ISO
  5. GB/T 18779.1-2022《产品几何技术规范(GPS) 工件与测量设备的测量检验 第1部分:按规范验证合格或不合格的判定规则》 · 国家市场监督管理总局、国家标准化管理委员会

How the guides are checked against these sources: About KELIBRON.

Watch it done

ReenactmentDigital caliper: outside diameter of Ground pin Ø12 · 0:28Watch in the lab → Follow step by step → Download video (12.0 MB) Video page →

Played back in the lab from domain commands; the reading and any verdict come from the measurement model.

Practise in the lab

Questions

What is the difference between error and uncertainty?

Error is a single difference from a reference value (often known only approximately); uncertainty quantifies the doubt about the result and is stated with a coverage factor.

Why divide by √3?

A limit ±a with every value equally likely (rectangular distribution) has a standard deviation of a/√3.