Tolerances and decisions
Measurement uncertainty basics
Error is the measured value minus a reference value (VIM 2.16); uncertainty is a non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used (VIM 2.26). A result is written as value ± U (k = 2, about 95 %).
On this page
Worked example: a caliper reading
| Source | Value | Standard uncertainty |
|---|---|---|
| Caliper MPE | ±0.02 mm | 0.02/√3 = 0.0115 mm |
| Resolution | 0.01 mm | 0.01/(2√3) = 0.0029 mm |
| Repeatability (10 readings) | s = 0.008 mm | 0.008/√10 = 0.0025 mm |
uc = √(0.0115² + 0.0029² + 0.0025²) ≈ 0.0122 mm (from the unrounded components), U = 2 × 0.0122 = 0.0244, rounded up to 0.025 mm. Mean reading 25.03 mm → 25.030 mm ± 0.025 mm (k = 2). Temperature, force and zero setting add further terms in a full budget. As an introductory budget the three terms above are treated as independent; a rigorous budget first checks for overlap, because calibration already covers part of what the MPE and repeatability terms describe.
Type A and Type B
- Type A: from repeated readings, using the standard deviation s of n readings; u = s/√n for the mean.
- Type B: from other information, such as a specification limit ±a with rectangular distribution gives u = a/√3; a resolution δ gives u = δ/(2√3).
- Combined: uc = √(u₁² + u₂² + …) for independent sources.
- Expanded: U = k · uc, usually k = 2.
Why it matters for pass/fail
If the tolerance band is 0.05 mm and U is 0.025 mm, half of the band is uncertainty, so the instrument is unsuitable (see the 10:1 rule). By the default rule of ISO 14253-1:2017 (when the supplier proves conformity), conformity is accepted only when the result lies inside the limits reduced on each side by a guard band of about 0.83U (1.65 combined standard uncertainties, clause 5.2.2, with U = 2u and a normal distribution whose standard deviation is clearly smaller than the tolerance; 95 % conformance probability); other decision rules can be agreed.
References
- JCGM 100:2008 — Evaluation of measurement data — Guide to the expression of uncertainty in measurement (GUM) · JCGM / BIPM
- JCGM 200:2012 — International vocabulary of metrology — Basic and general concepts and associated terms (VIM, 3rd edition) · JCGM / BIPM
- JJF 1059.1-2012《测量不确定度评定与表示》 (cited by number; no public page) · 国家质量监督检验检疫总局
- ISO 14253-1:2017 — GPS — Inspection by measurement of workpieces and measuring equipment — Part 1: Decision rules for verifying conformity or nonconformity with specifications · ISO
- GB/T 18779.1-2022《产品几何技术规范(GPS) 工件与测量设备的测量检验 第1部分:按规范验证合格或不合格的判定规则》 · 国家市场监督管理总局、国家标准化管理委员会
How the guides are checked against these sources: About KELIBRON.
Watch it done
Played back in the lab from domain commands; the reading and any verdict come from the measurement model.
Practise in the lab
Questions
What is the difference between error and uncertainty?
Error is a single difference from a reference value (often known only approximately); uncertainty quantifies the doubt about the result and is stated with a coverage factor.
Why divide by √3?
A limit ±a with every value equally likely (rectangular distribution) has a standard deviation of a/√3.
Try it yourself
Open the same setup in the 3D lab and take the reading yourself.
Related guides
Tolerances and decisionsTolerances and fits basicsA toleranced size gives a nominal size and two deviations.
Tolerances and decisionsAcceptance limits and the safety margin AUnder GB/T 3177-2009 a plain size is accepted by acceptance limits.
Tolerances and decisionsAn uncertainty budget for a caliper measurementAn uncertainty budget lists every source of uncertainty, turns each into a standard uncertainty, combines them as the square root of the sum of squares and multiplies by a coverage factor.