Checking a caliper and a micrometer with gauge blocks

3 min readUpdated

Key answer

Put a gauge block between the measuring faces, read the instrument and subtract the block’s calibrated size: the difference is the error at that point. Do this at five or more roughly equally spaced points over a caliper’s range, and with a set of blocks chosen to show a micrometer screw’s errors. Comparing each error with the maximum permissible error and recording pass or fail is verification; a calibration reports the errors with their uncertainty.

On this page
  1. Caliper: outside jaws
  2. Parallelism and the inside jaws
  3. Micrometer
  4. Recording and deciding
  5. Planned for the lab
  6. References
  7. Practise the shared basics on a caliper
  8. Questions

Caliper: outside jaws

  1. Let the caliper and the blocks reach the same temperature on the bench; handle the blocks with gloves or tongs, not bare hands.
  2. Close the jaws and check zero (daily care and the zero check are in caliper care and calibration).
  3. Put a block between the outside measuring faces and read. Error = reading − calibrated size of the block.
  4. Repeat at five or more roughly equally spaced points covering the range of the main scale and the vernier (NPL GPG 40).

For a 0–150 mm digital caliper, 30, 60, 90, 120 and 150 mm would do; this is the example caliper used in how to read a calibration certificate. On a vernier caliper, also choose sizes whose decimal part falls on different vernier lines, so the vernier scale is checked as well as the main scale.

Parallelism and the inside jaws

Put a block at different points along the jaw, near the root and near the tip. Use two blocks of different length so the check is made at two places in the range, one close to zero and one close to full range, and check that clamping the slider does not change the result.

For the inside jaws, build a known internal size from a gauge block with a holder and accessory jaws, measure it with the inside faces and subtract the block size in the same way.

NPL also checks the squareness of the fixed face to the guiding edge with a knife-edge square, the flatness of the faces with an optical flat and their straightness with a lever-type dial indicator.

Micrometer

A micrometer is checked with a special set of gauge blocks whose sizes are chosen to show both the periodic error of the screw (within a turn) and its progressive error (along the travel), and with optical parallels for the flatness and parallelism of the faces. Put each block in turn between the measuring faces (anvil and spindle), using the ratchet as when measuring, and note the reading. A micrometer stand keeps your hand off the frame.

NPL’s good practice guide gives this example of a micrometer calibration (positions from 2.5 to 25 mm):

Position (mm)Error (mm)
2.50.000
10.3+0.001
12.9−0.001
15−0.002
20.2−0.002
25−0.003

Its expanded uncertainty is ±0.002 mm (k = 2, about 95 %). The same guide notes that most organisations send their micrometers to an accredited laboratory rather than calibrating them in house.

Recording and deciding

Record the block set and its certificate number, the temperature, each reading and error, the maximum permissible error used, the decision, who checked and the date (see gauge management). Comparing each error with the MPE and writing pass or fail is verification (in China, calipers are verified to JJG 30 and micrometers to JJG 21). A calibration certificate from a laboratory reports the errors and their uncertainty instead.

Planned for the lab

Gauge blocks of 10, 20, 25, 30 and 40 mm are in KELIBRON as setting standards for the dial indicator. A micrometer can be closed on them, but the lab’s calipers and micrometers have no error of indication that varies over the range yet (only the zero-error demonstration, a constant offset), so a block check would find nothing to record. Until that is added, practise the habits on the scenes below: check zero first and read to the instrument’s resolution.

References

  1. D. Flack, Callipers and Micrometers, NPL Measurement Good Practice Guide No. 40, Issue 2, 2014 (first issued 2001) · National Physical Laboratory
  2. JCGM 200:2012 — International vocabulary of metrology — Basic and general concepts and associated terms (VIM, 3rd edition) · JCGM / BIPM
  3. JJG 30-2012《通用卡尺》检定规程 (cited by number; no public page) · 国家质量监督检验检疫总局
  4. JJG 21-2008《千分尺检定规程》 (cited by number; no public page) · 国家质量监督检验检疫总局
  5. ISO 13385-1:2019 — Geometrical product specifications (GPS) — Dimensional measuring equipment — Part 1: Design and metrological characteristics of callipers · ISO
  6. ISO 3611:2023 — Geometrical product specifications (GPS) — Dimensional measuring equipment — Design and metrological characteristics of micrometers for external measurements (cited by number; no public page) · ISO

How the guides are checked against these sources: About KELIBRON.

Practise the shared basics on a caliper

Questions

How many points should I check on a caliper?

At least five, roughly equally spaced, covering the range (NPL GPG 40), plus a parallelism check near zero and near full range.

Is checking against gauge blocks a calibration?

Finding the errors against traceable blocks is the measuring part of a calibration; a calibration also states their uncertainty. Deciding pass or fail against the MPE is verification.

Can I calibrate a micrometer myself?

You can check it against a suitable block set; NPL notes that most organisations send micrometers to an accredited laboratory.

Practise the basics first

Cleaning, zero checks, square positioning, light even force and reading to the scale interval are the same on every instrument; practise them in a caliper scene.